Reference Manual Rev 1.13
004R-646-113
Page 63
12.3. SCALE (Loadcell options and calibration)
12.3.1.
BUILD (Scale parameters)
Path
Description
SCALE
└ BUILD
└
TYPE
(
)
└
CABLE
(
)
└
DP
(
)
└
CAP1
(
)
└
E1
(
)
└
CAP2
(*
)
└
E2
(*
)
└
UNITS
(
)
└
HI.RES
(
)
Scale Base configuration settings:
TYPE
: Range type. Options are:
SINGLE : Single range
DUAL.I: Dual interval
DUAL.R: Dual range
CABLE
: 6-wire or 4-wire cable termination:
6-wire: SENSE lines are connected to the
instrument.
4-wire: Internal connection between
Excitation and SENSE lines is active.
DP:
Set the decimal point position.
CAP1:
Sets the fullscale capacity for the scale. If
using multiple interval/range, this sets the fullscale
capacity of the lowest range/interval.
E1
: Sets the count-by (or resolution) of the scale. If
using multiple interval/range, this sets the count-by
(or resolution) of the lowest range/interval.
CAP2
: If using multiple interval/range, this sets the
fullscale capacity of the highest range/interval.
E2
: If using multiple interval/range, this sets the
count-by (or resolution) of the highest range/interval.
UNITS
: Sets the weighing units.
NB:
For Options:
None: Units are left blank.
ARROW.U: Use the top arrow. Units
will be printed onto the instrument in the
correct location.
HI.RES
: Sets the scale to high resolution (x10)
mode.
: This item is trade critical and will affect the
calibration counter(s) if changed.
TYPE Values
(
)
<OPT>
SINGLE
(Default)
DUAL.I , DUAL.R
CABLE Values
(
)
<OPT>
6 WIRE
(Default)
, 4 WIRE
DP Values
(
)
<OPT>
000000
(Default)
00000.0
0000.00
000.000
00.0000
0.00000
CAP1
&
CAP2 Values
(
)
<NUM>
100 ..999999
Default: 3000
NB:
Numbers above assume
no decimal point.
E1 & E2 Values
(
)
<OPT>
1
(Default)
2
5
10
20
50
100
UNITS Values
(
)
<OPT>
None
kg
(Default)
lb
t
g
Oz
N
ARROW U
HI.RES Values
(
)
<OPT>
OFF
(Default)
, ON