P
AGE
6-16
T
RILOGY
S
ERVICE
& T
ECHNICAL
I
NFORMATION
1002735, V
ER
. 05
High
Temperature
Escalates
from
Medium to
High
• • •
for
medium
• • • • •
for high
It occurs when the
estimated patient
airstream temperature
or the ventilator internal
temperature is too high.
The alarm occurs in
several stages.
Internal fans are
started when the
medium priority alarm
is generated. If the
condition causing the
high temperature is not
corrected and the
temperature continues
to rise, the alarm will
escalate to the high
priority alarm.
Yellow flashing
button and yellow
“High
Temperature”
message
appears. If
condition
worsens, button
flashes red and
red “High
Temperature”
message
appears.
Operates
Replace
Detachable
Battery
Low or
High,
depending
on cause
of alarm
• •
for low
• • • • •
for high
It occurs when the
detachable battery is
nearing the end of its
useful life or a failure in
the detachable battery
that prevents it from
charging or discharging
has been detected.
The alarm occurs in
several stages, from
low to high priority. If
the alarm is reset
without removing the
battery, the alarm will
be regenerated once
every hour until the
detachable battery is
removed. The
detachable battery is
not used, and the
power source is
switched to the next
available power source
if the alarm is the high
priority alarm.
“Replace
Detachable
Battery” message
appears. If
battery is nearing
end of useful life,
message appears
with yellow
background and
button is solid
yellow. If battery
fails, message
appears with red
background and
button flashes red.
Operates
A
LARM
P
RIORITY
A
UDIBLE
D
ESCRIPTION
OF
A
LARM
V
ISUAL
(A
LARM
I
NDICATOR
B
UTTON
AND
D
ISPLAY
)
D
EVICE
A
CTION
Summary of Contents for Trilogy 02
Page 1: ...Trilogy 100 Trilogy 200 Trilogy 02 Trilogy 202 Service Technical Information ...
Page 177: ...PAGE 8 16 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 3 4 5 ...
Page 181: ...PAGE 8 20 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 6 ...
Page 184: ...PAGE 8 23 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 343: ...PAGE 10 40 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 346: ...PAGE 10 43 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 351: ...PAGE 10 48 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 352: ...PAGE 10 49 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 353: ...PAGE 10 50 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 354: ...PAGE 10 51 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 418: ...PAGE 12 5 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 419: ...PAGE 12 6 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 420: ...PAGE 12 7 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 421: ...PAGE 12 8 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 422: ...PAGE 12 9 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 423: ...PAGE 12 10 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...