P
AGE
6-50
T
RILOGY
S
ERVICE
& T
ECHNICAL
I
NFORMATION
1002735, V
ER
. 05
E-184
Low Detachable
Battery
•
Low battery - Detachable Li-
Ion has < 20 minutes run time
remaining and it is last
available power source
•
Charge Detachable Li-Ion
Battery
•
Replace Detachable Li-Ion
Battery
E-185
Low Internal
Battery
•
Low battery - Internal Li-Ion
has < 20 minutes run time
remaining and it is last
available power source
•
Charge Internal Li-Ion Battery
E-186
Low External
Battery
•
Depleted battery - Lead Acid
has < 10 minutes run time
remaining and it is last
available power source
•
Charge Lead Acid battery
•
Replace Lead Acid battery
E-187
Low Detachable
Battery
•
Depleted battery - Detachable
Li-Ion has < 10 minutes run
time remaining and it is last
available power source
•
Charge Detachable Li-Ion
Battery
•
Replace Detachable Li-Ion
battery
E-188
Low Detachable
Battery
•
Depleted battery - Internal Li-
Ion has < 10 minutes run time
remaining and it is last
available power source
•
Charge Internal Li-Ion Battery
E-192
Replace
Detachable
Battery
•
AFE chip does not receive the
appropriate frequency on the
WDI pin from the Gas Gauge
chip. Both of these chips are
in the Li-Ion battery pack
•
Detachable Li-Ion fault
•
Replace Detachable Li-Ion
battery
•
Replace Power Management
Board
E-194
Replace
Detachable
Battery
•
Detachable Li-Ion Battery
Permanent Failure
•
Replace Detachable Li-Ion
battery
E-196
Replace
Detachable
Battery
•
Detachable Li-Ion State of
Health < 50%. Full Charge
Capacity (FCC) is less than
51% of the design capacity
•
Detachable Li-Ion FCC error
•
Detachable Li-Ion fault
•
Detachable Li-Ion end of life
•
Cycle the Detachable Li-Ion
(discharge battery so that
capacity change is > 50%, then
rest for > 2 hours. Charge to
100%, then rest for > 2 hours)
•
Replace Detachable Li-Ion
battery
E-206
Replace
Detachable
Battery
•
Corruption of battery gas
gauge
•
Replace Battery
•
Cycle the Detachable Li-Ion
(diecharge battery so that
capacity change is > 50%, then
rest for > 2 hours. Charge to
100%, then rest for > 2 hours)
Summary of Contents for Trilogy 02
Page 1: ...Trilogy 100 Trilogy 200 Trilogy 02 Trilogy 202 Service Technical Information ...
Page 177: ...PAGE 8 16 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 3 4 5 ...
Page 181: ...PAGE 8 20 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 6 ...
Page 184: ...PAGE 8 23 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 343: ...PAGE 10 40 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 346: ...PAGE 10 43 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 351: ...PAGE 10 48 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 352: ...PAGE 10 49 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 353: ...PAGE 10 50 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 354: ...PAGE 10 51 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 418: ...PAGE 12 5 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 419: ...PAGE 12 6 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 420: ...PAGE 12 7 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 421: ...PAGE 12 8 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 422: ...PAGE 12 9 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 423: ...PAGE 12 10 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...