P
AGE
7-21
T
RILOGY
S
ERVICE
& T
ECHNICAL
I
NFORMATION
1002735, V
ER
. 05
7.6 I
NTERNAL
B
ATTERY
C
ONDITIONING
P
ROCEDURE
Respironics recommends replacement of the Internal Battery when the battery reaches 400 or more cycles. If
the Internal Battery is not replaced and the Cycle Count is greater than 400 or the Max Error is greater than
25% then the following procedure must be performed to ensure it is still functioning properly.
1. Remove the Detachable Battery.
2. Set the device to the following therapy settings:
•
T mode
•
IPAP = 34 pressure units
•
EPAP = 4 pressure units
•
BPM = 10
•
Rise Time = 1
•
Inspiratory Time = 1 second
3. Connect AC, turn on Ventilator and charge battery to 100% capacity. Battery is fully charged when
the Lightning bolt charge indicator is not displayed.
4. Disconnect AC and allow battery to discharge until Ventilator turns off.
5. Allow battery to rest for a minimum of 5.5 hours. Do not connect AC power and do not attempt to
turn the Ventilator on.
6. Connect AC power and allow the Internal Battery to charge to 100%. Battery is fully charged when
the Lightning bolt charge indicator is not displayed.
7. Allow battery to rest for minimum 2.5 hours. Do not disconnect AC power during this time.
8. Open the Trilogy Toolbox software and read the Internal Battery State of Health and MaxError
using the Tool Box. Refer to Section 7.5 for more instructions.
9. The Internal Battery must have a State of Health > 60% and MaxError =1%. If the Internal Battery
does not meet the pass criteria it must be replaced.
NOTE
If the Trilogy Toolbox software is not downloaded to your computer, proceed to the Testing
Chapter of this manual for download instructions.
Summary of Contents for Trilogy 02
Page 1: ...Trilogy 100 Trilogy 200 Trilogy 02 Trilogy 202 Service Technical Information ...
Page 177: ...PAGE 8 16 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 3 4 5 ...
Page 181: ...PAGE 8 20 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 6 ...
Page 184: ...PAGE 8 23 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 343: ...PAGE 10 40 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 346: ...PAGE 10 43 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 351: ...PAGE 10 48 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 352: ...PAGE 10 49 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 353: ...PAGE 10 50 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 354: ...PAGE 10 51 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 418: ...PAGE 12 5 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 419: ...PAGE 12 6 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 420: ...PAGE 12 7 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 421: ...PAGE 12 8 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 422: ...PAGE 12 9 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 423: ...PAGE 12 10 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...