P
AGE
10-66
T
RILOGY
S
ERVICE
& T
ECHNICAL
I
NFORMATION
1002735, V
ER
. 05
10.10 T
RILOGY
F
IELD
S
ERVICE
A
PPLICATION
This application must be performed after repair of a Trilogy Ventilator or during routine maintenance as
specified in the Maintenance Section of this Service Manual.
10.10.1E
QUIPMENT
R
EQUIRED
1. PC System with 1 Serial Port and at least 6 USB Ports with Windows XP installed
2. Differential Pressure Indicator (Required only for Trilogy 200, Trilogy O
2
, & Trilogy 202)
3. Digital Manometer (0-70 PSI Pressure Meter for Trilogy 100 & Trilogy 200 / 0-100 PSI Pressure
Meter for Trilogy O
2
& Trilogy 202)
4. External Power Supply Capable of providing 15 VDC and 24 Amps or Deep Cycle Marine Battery
12V
5. TSI Model 4040 Flow Meter
6. Flow Control Valve
7. USB to Serial Converter
8. Temperature and RH Meter
9. Trilogy to PC Data Cable (Quantity of 2 for Auto Flow testing)
10. Serial RS-232 Cables (Quantity 3)
11. Merriam Pressure Pump w/ Vernier (Required only for Trilogy 200, Trilogy O
2
, & Trilogy 202)
12. Air Filtration and Regulation Assembly
13. Digital Multimeter capable of providing True RMS Measurements
14. Ethernet CAT 5 or 6 Cable
15. Linksys Network Switch Minimum 2 port
16. Test Orifice, .25” ID
17. Trilogy Nurse Call Adapter Cable
18. Outlet Port Cap (Quantity 2)
19. Smoothbore Tubing, 18” (Quantity 2)
20. SD Card
21. Exhalation Porting Block, Universal
22. Exhalation Porting Block, Passive
23. Trilogy Test Hardware Kit
24. O
2
Vent Ports (Quantity 4)
25. Smoothbore Tubing, 6 ft.
26. Brass Barb Fitting Male for 1” hose ID and 3/4” Pipe
27. Reducing Hex Coupler 3/4” to 1/2” Female to Female
NOTE
Prior to performing any of the testing sections, please allow equipment a warm-up period of no less than
10 minutes for stabilization. Perform self-cal on the RI Manometer and Druck DPI-150 if necessary.
If
ONLY
the AC Connector has been replaced the Performance Verification Tool detailed in the
Maintenance Section of this manual may be used in place of the Field Service Application.
Summary of Contents for Trilogy 02
Page 1: ...Trilogy 100 Trilogy 200 Trilogy 02 Trilogy 202 Service Technical Information ...
Page 177: ...PAGE 8 16 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 3 4 5 ...
Page 181: ...PAGE 8 20 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 6 ...
Page 184: ...PAGE 8 23 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 343: ...PAGE 10 40 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 346: ...PAGE 10 43 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 351: ...PAGE 10 48 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 352: ...PAGE 10 49 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 353: ...PAGE 10 50 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 354: ...PAGE 10 51 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 418: ...PAGE 12 5 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 419: ...PAGE 12 6 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 420: ...PAGE 12 7 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 421: ...PAGE 12 8 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 422: ...PAGE 12 9 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 423: ...PAGE 12 10 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...