P
AGE
7-1
T
RILOGY
S
ERVICE
& T
ECHNICAL
I
NFORMATION
1002735, V
ER
. 05
C
HAPTER
7: M
AINTENANCE
7.0 C
HAPTER
O
VERVIEW
This chapter describes the Maintenance intervals and procedures for the Trilogy Ventilator.
7.1 R
OUTINE
M
AINTENANCE
Routine maintenance involves periodic checking, cleaning, and or replacing the following item as necessary:
•
Air Inlet Filter
•
Enclosures
7.1.1 A
IR
I
NLET
F
ILTER
Under normal usage, you should clean the inlet filter at least once every two weeks and replace it with a new
filter every six months or sooner if needed.
7.1.2 E
NCLOSURES
Commensurate with hospital or homecare policies, Respironics recommends cleaning the cabinet and
inspecting for damage as necessary.
7.2 P
REVENTIVE
M
AINTENANCE
The following Preventive Service procedures will be followed by authorized service personnel at the intervals
outlined in the maintenance schedule below. Preventive Maintenance must be performed by Respironics or a
Respironics’ trained location.
Your new Philips Respironics Trilogy ventilator includes a PM Service Due label affixed to the bottom of your
device. The Service Date line on the label is blank, which allows you to document the date when service is
due. The Service Hours line will have 10000 hours recorded. Service is due every 10000 blower hours or 24
months whichever comes first depending on the usage of the device or 17,500 blower hours for blower
maintenance.
During the initial install, you will need to record the date and hours (if not already written) on the blank label to
maximize your in-service interval. Record the service date as 24 months from the date first placed on a
patient. Record the hours as 10000 blower hours for initial usage.
The device blower hours are listed in the device software in the information menu.
A new blank label should be applied to the Trilogy Ventilator after the 10000 hours or 24 months maintenance
and after the 17,500 Blower Maintenance has been performed. The label should be completed according to
the table below with the new next maintenance due blower hours and new next maintenance due date (should
be 24 months from date of completed Maintenance) before returning the ventilator.
Summary of Contents for Trilogy 02
Page 1: ...Trilogy 100 Trilogy 200 Trilogy 02 Trilogy 202 Service Technical Information ...
Page 177: ...PAGE 8 16 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 3 4 5 ...
Page 181: ...PAGE 8 20 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 6 ...
Page 184: ...PAGE 8 23 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 343: ...PAGE 10 40 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 346: ...PAGE 10 43 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 351: ...PAGE 10 48 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 352: ...PAGE 10 49 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 353: ...PAGE 10 50 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 354: ...PAGE 10 51 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 418: ...PAGE 12 5 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 419: ...PAGE 12 6 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 420: ...PAGE 12 7 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 421: ...PAGE 12 8 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 422: ...PAGE 12 9 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 423: ...PAGE 12 10 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...