P
AGE
10-32
T
RILOGY
S
ERVICE
& T
ECHNICAL
I
NFORMATION
1002735, V
ER
. 05
Verify Reset
1. Reconnect or turn on the source of high pressure O
2
to the ventilator.
2. Wait 1 minute and verify the following:
•
The set level of FiO
2
is satisfied using an external O
2
monitor.
•
The High Priority audible indicator has stopped sounding.
•
The red light on the Alarm Indicator/Audio Pause button has stopped flashing.
10.5.5 B
ATTERY
F
UNCTION
V
ERIFICATION
Make sure the batteries are functioning properly and fully charged before patient use.
V
ERIFY
THE
D
ETACHABLE
AND
I
NTERNAL
(L
ITHIUM
-I
ON
) B
ATTERIES
F
UNCTION
1. Connect AC Power to the device and verify that the green AC LED on the front panel is lit.
2. Verify that the detachable battery is properly installed.
3. Turn the device on and verify that both the detachable and internal battery symbols appear on the
display. Verify that if either battery is less than fully charged, the charge symbol will display on the
respective battery.
4. Disconnect the AC Power source from the device.
•
Verify that the AC Power Disconnected alarm message appears on the display and the green
AC LED is not lit. Press Reset.
•
Verify that the detachable battery symbol shows the level of charge noted in the previous step
and that the device continues to operate.
•
Verify that the detachable battery symbol has a black box around it to indicate that it is in use.
5. Disconnect the detachable battery pack from the device.
•
Verify that the Detach Batt Disconnected alarm message appears on the display. Press Reset.
•
Verify that the internal battery symbol shows the same level of charge as noted in Step C and
the device continues to operate.
•
Verify that the internal battery symbol has a black box around it to indicate that it is in use.
6. Reconnect the Detachable Battery and AC Power source.
V
ERIFY
THE
E
XTERNAL
B
ATTERY
F
UNCTION
(I
F
A
VAILABLE
)
1. Connect AC Power to the device and verify that the green AC LED is lit.
2. Connect the external battery cable to the external battery and to the ventilator.
3. Verify that the external battery symbol is shown on the display and some level of charge is present.
4. Disconnect the AC Power source from the device.
•
Verify that the AC Power Disconnected alarm message appears on the display and the green
AC LED is not lit. Press Reset.
•
Verify that the external battery symbol shows the level of charge as noted in the previous step
and the device continues to operate.
•
Verify that the external battery symbol has a black ••box around it to indicate that it is in use.
5. Reconnect the AC Power source.
A
LARM
AND
E
VENT
L
OG
C
LEAN
-U
P
1. In the Setup Menu, select Alarm Log.
Summary of Contents for Trilogy 02
Page 1: ...Trilogy 100 Trilogy 200 Trilogy 02 Trilogy 202 Service Technical Information ...
Page 177: ...PAGE 8 16 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 3 4 5 ...
Page 181: ...PAGE 8 20 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 6 ...
Page 184: ...PAGE 8 23 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 343: ...PAGE 10 40 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 346: ...PAGE 10 43 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 351: ...PAGE 10 48 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 352: ...PAGE 10 49 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 353: ...PAGE 10 50 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 354: ...PAGE 10 51 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 418: ...PAGE 12 5 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 419: ...PAGE 12 6 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 420: ...PAGE 12 7 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 421: ...PAGE 12 8 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 422: ...PAGE 12 9 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 423: ...PAGE 12 10 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...