P
AGE
6-4
T
RILOGY
S
ERVICE
& T
ECHNICAL
I
NFORMATION
1002735, V
ER
. 05
If a Ventilator Inoperative alarm occurs, the entire display screen turns red and the Ventilator inoperative
message appears on-screen, as shown below.
To turn the ventilator off from a Ventilator Inoperative condition, use the normal power off sequence. When the
Start/Stop button is selected, the following screen will display.
Select the Right button (Yes) to turn the ventilator off and stop the audible alarm. Selecting the Left button (No)
will return the screen to the Ventilator Inoperative Alarm Screen without silencing the audible alarm.
6.2.4 R
EMOTE
A
LARM
When the ventilator detects an alarm condition, if you are using a remote alarm system, a signal is sent from
the ventilator to activate the remote alarm.
6.3 A
UDIO
P
AUSE
AND
A
LARM
R
ESET
F
EATURES
This section describes the Audio Pause and Alarm Reset features.
6.3.1 A
UDIO
P
AUSE
When an alarm occurs, you can temporarily silence the audible indicator by pressing the Alarm Indicator/Audio
Pause button. The alarm is silenced for 60 seconds and then will sound again if the cause of the alarm has not
been corrected. Each time the Alarm Indicator/Audio Pause button is pressed, the alarm silence period resets
to one minute.
Summary of Contents for Trilogy 02
Page 1: ...Trilogy 100 Trilogy 200 Trilogy 02 Trilogy 202 Service Technical Information ...
Page 177: ...PAGE 8 16 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 3 4 5 ...
Page 181: ...PAGE 8 20 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 6 ...
Page 184: ...PAGE 8 23 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 343: ...PAGE 10 40 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 346: ...PAGE 10 43 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 351: ...PAGE 10 48 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 352: ...PAGE 10 49 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 353: ...PAGE 10 50 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 354: ...PAGE 10 51 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 418: ...PAGE 12 5 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 419: ...PAGE 12 6 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 420: ...PAGE 12 7 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 421: ...PAGE 12 8 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 422: ...PAGE 12 9 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 423: ...PAGE 12 10 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...