Turn Device Power On
Press the Start/Stop button on the front of the ventilator. The system will begin operating using the defined
ventilation settings.
V
ERIFY
THE
H
IGH
T
IDAL
V
OLUME
A
LARM
This procedure verifies that the High Tidal Volume alarm is working properly. For Passive and Active Flow
circuits, this will verify the High Vte alarm. For Active PAP circuits, this will verify the High Vti alarm. It assumes
that you have attached the test lung, verified the ventilator settings, and turned on ventilator power as
described in the Initial Setup section.
Change Alarm Ventilator Setting
Modify the High Tidal Volume alarm setting to match the one shown below.
Verify the Alarm
Wait up to 40 seconds and verify the following alarm signals:
•
The High Priority audible indicator sounds
•
A red light flashes on the Alarm Indicator/Audio Pause button
•
The High Tidal Volume alarm condition appears on the screen, highlighted in red
Modify Ventilator Alarm Settings
Modify the High Tidal Volume alarm setting to match the one shown below.
Verify Reset
Wait 40 seconds and verify the following auto-reset conditions:
•
The High Priority audible indicator has stopped sounding
•
The red light on the Alarm Indicator/Audio Pause button has stopped flashing
NOTE
Do not use the “reset” button to manually reset the alarm. Instead, use the “Modify”
button to change ventilator settings. This note applies to all tests.
SETTING
VALUE
High Vte/High Vti
50 ml
SETTING
VALUE
High Vte/High Vti
500 ml
Summary of Contents for Trilogy 02
Page 1: ...Trilogy 100 Trilogy 200 Trilogy 02 Trilogy 202 Service Technical Information ...
Page 177: ...PAGE 8 16 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 3 4 5 ...
Page 181: ...PAGE 8 20 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 6 ...
Page 184: ...PAGE 8 23 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 343: ...PAGE 10 40 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 346: ...PAGE 10 43 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 351: ...PAGE 10 48 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 352: ...PAGE 10 49 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 353: ...PAGE 10 50 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 354: ...PAGE 10 51 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 418: ...PAGE 12 5 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 419: ...PAGE 12 6 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 420: ...PAGE 12 7 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 421: ...PAGE 12 8 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 422: ...PAGE 12 9 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 423: ...PAGE 12 10 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...