P
AGE
6-38
T
RILOGY
S
ERVICE
& T
ECHNICAL
I
NFORMATION
1002735, V
ER
. 05
E-267
•
FCC > 150% of the Li-Ion design capacity
•
Corrupted Gas Gauge
•
Cycle the Internal Li-Ion (discharge
battery so that capacity change is
> 50%, then rest for > 2 hours.
Charge to 100%, then rest for > 2
hours)
•
Replace Internal Li-Ion battery
E-269
•
Error indicates an open VBULK FET. This
MOSFET connects the output of the AC/DC
supply to the input of the boost converter. Unit
will not run on AC power. Will use battery power
if available
•
Replace Power Management PCA
E-270
•
Error indicates an open Q11, which controls the
discharge of the Li-Ion batteries post VbattLi.
When the Lithium Ion power is selected, this FET
is turned on. If Q11 fails open, there will be no Li-
Ion battery power to the system
•
Replace Power Management PCA
E-272
•
Charger chip failure
•
FET failure
•
Replace Power Management PCA
E-273
•
Unit not calibrated
•
Calibration table corrupted
•
Recalibrate the device
•
Replace the System/CPU PCA
E-274
•
Unit not calibrated
•
Calibration table corrupted
•
Recalibrate the device
•
Replace the System/CPU PCA
E-290
•
One or more “Ventilator Service Required” errors
are active in the system
•
Examine the Significant Event Log
error log. The first column of data
contains the first error code that is
causing this error message.
•
Address according to this first error
code
•
Each time you address an error,
cycle the motor (OFF to ON to
OFF) and check the Significant
Event Log again until this alarm is
no longer sounded
•
Remove all power from the unit
and then reapply power, then turn
the motor ON to see if this
message is still reported
E-302
•
Voltage reference on internal ADC failed
•
Replace System/CPU PCA
E-311
•
Unit wasn’t calibrated or calibration table was
destroyed
•
Recalibrate
•
Replace System/CPU PCA
Summary of Contents for Trilogy 02
Page 1: ...Trilogy 100 Trilogy 200 Trilogy 02 Trilogy 202 Service Technical Information ...
Page 177: ...PAGE 8 16 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 3 4 5 ...
Page 181: ...PAGE 8 20 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 6 ...
Page 184: ...PAGE 8 23 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 343: ...PAGE 10 40 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 346: ...PAGE 10 43 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 351: ...PAGE 10 48 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 352: ...PAGE 10 49 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 353: ...PAGE 10 50 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 354: ...PAGE 10 51 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 418: ...PAGE 12 5 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 419: ...PAGE 12 6 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 420: ...PAGE 12 7 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 421: ...PAGE 12 8 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 422: ...PAGE 12 9 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 423: ...PAGE 12 10 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...