P
AGE
8-70
T
RILOGY
S
ERVICE
& T
ECHNICAL
I
NFORMATION
1002735, V
ER
. 05
8.6 F
RONT
/R
EAR
/B
OTTOM
E
NCLOSURE
I
NSTALLATION
1.
IMPORTANT TO BE CONNECTED LAST:
Connect the Internal Battery Cable to J2 on the Power
Management Board
2. Place the Top Plate in the groves on the Front and Rear Enclosure.
3. Move the enclosure close enough to connect the wiring.
4. Using two tie wraps, secure the Ferrite on the Blower wires to the plastic support bracket.
5. Connect the Detachable Battery cable to J1 and J3 on the Power Management Board.
6. Connect the Exhaust Fan to J1 on the System Board.
7. Connect the Exhalation Control Module to J10 on the System Board.
8. Secure the Exhalation Control Module Wires to the Metal Support using cloth tape.
9. Connect the Blower Fan to J17 on the System Board.
10. Connect the Blower to J5 on the System Board.
11. Connect the Sensor Cable to J4 on the Sensor Board.
12. Ensuring not to pinch tubing or wiring, slide the Front and Rear Enclosures completely together.
13. Secure the Front and Rear Enclosures together by tightening the screws to 8 in-lbs.
14. The Front and Rear Enclosure are now in place. It is now time to connect the Bottom Enclosure.
15. Connect the Capacitor to J6 on the System Board.
16. Connect the Battery Fan Cable to J18 on the System Board.
17. Connect the AC Connector to J1 on the Power Supply Board.
18. Connect the Power Supply to location J1 on the Power Management Board.
19. Connect the Interface Cable to J7 on the System Board.
20. Connect the Ethernet Cable to J19 on the System Board.
21. Connect the Oxygen Tube to the Exhalation Control Module.
22. Connect the DC Battery Cable to J4 on the Power Management Board.
23. Connect the OBM wire harness to J7 on the Interface PCA (Trilogy O
2
and Trilogy 202 Only).
24.
IMPORTANT TO BE CONNECTED LAST:
Connect the Internal Battery Cable to J2 on the Power
Management Board.
25. Ensure all wiring connections are still connected and the Oxygen Tube is still connected.
26. Fold the Bottom Enclosure over to seal the enclosure together. Ensure that the wiring and tubing
is clear of the seam and not kinked.
27. Lift up the device and turn the device to have the LCD face you. Gently lift up one end of the Bot-
tom Enclosure and connect the Alarms to Speaker 1 and Speaker 2 on the front Panel Board.
Ensure that the wiring and tubing is clear of the seam and not kinked.
28. Secure the Bottom Enclosure to the Front and Rear Enclosures by tightening the screws to 8 in-
lbs.
29. Connect the Active/Passive Port. Refer to the Active/Passive Porting Block Replacement Section
30. Insert the Air Path Foam. Refer to the Air Path Foam Replacement Section for more details.
CAUTION
Once the Internal Battery Cable is connected, power may be active.
Summary of Contents for Trilogy 02
Page 1: ...Trilogy 100 Trilogy 200 Trilogy 02 Trilogy 202 Service Technical Information ...
Page 177: ...PAGE 8 16 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 3 4 5 ...
Page 181: ...PAGE 8 20 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 6 ...
Page 184: ...PAGE 8 23 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 343: ...PAGE 10 40 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 346: ...PAGE 10 43 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 351: ...PAGE 10 48 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 352: ...PAGE 10 49 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 353: ...PAGE 10 50 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 354: ...PAGE 10 51 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 418: ...PAGE 12 5 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 419: ...PAGE 12 6 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 420: ...PAGE 12 7 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 421: ...PAGE 12 8 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 422: ...PAGE 12 9 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Page 423: ...PAGE 12 10 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...