Manual - TC.P Quadro
Commissioning
2016-08-05
50/330
V04.52
3
Electromagnetic compatibility
TopCon power supplies have interference protection and suppression
filters on all power and signal connections such that immunity is
achieved on correct installation compliance with the applicable IEC and
EN standards in relation to interference.
The following standards are applicable:
Interference immunity:
EN 61000-6-2
Interference emission:
EN 61000-6-4
For the interference protection components to be able to perform their
function, the following conditions must apply:
Large area earthing suitable for EMI.
Shield mains and load connection.
(Depending on load and installation situation)
Connect both ends of shields to earth.
(Depending on load and installation situation)
Interference immunity
The system must be fully earthed (encompassing all parts) and earthed
suitable for EMI.
Correct earthing and shielding of the cable connections are key ele-
ments for interference immunity.
All cable shields must be connected to earth potential at both
ends using large area connections.
Ideally shielded connectors are used that are earthed directly via
the earthed sockets on the electrical power supply.
Star topology earthing (suitable for EMI) of the device.