PULSAR MEASUREMENT
71
ENCLOSURE INSTALLATION
Locate the enclosure within 25 ft (7.6 m) of the sensors (up to 100 ft - 30 m optional). The enclosure can be wall
mounted with the four mounting screws (included) or panel mounted with Option PM Panel Mount kit from Pulsar
Measurement.
Avoid mounting the enclosure in direct sunlight to protect the electronics from damage due to overheating and
condensate. In high humidity atmospheres, or where temperatures fall below freezing, Option TH Enclosure
Heater and Thermostat is recommended. IMPORTANT: Seal conduit entries to prevent moisture from entering
enclosure.
NEMA4X (IP66) WITH CLEAR COVER
1. Open hinged enclosure cover.
2. Insert #12 screws (supplied) through the four enclosure mounting holes to
secure the enclosure to the wall or mounting stand.
Additional conduit holes can be cut in the bottom of the enclosure when required.
Use a hole saw or Greenlee-type hole cutter to cut the required holes.
IMPORTANT:
DO NOT make conduit/wiring entries into the top or sides of the
enclosure.
Note: This non-metallic enclosure does not automatically provide
grounding between conduit connections. Grounding must be provided as
part of the installation. Ground in accordance with the requirements of the
National Electrical Code. System grounding is provided by connecting
grounding wires from all conduit entries to the steel mounting plate or
another point which provides continuity.
CLEANING
Cleaning is not required as a part of normal maintenance.