PULSAR MEASUREMENT
25
SPECIAL FUNCTIONS (cont.)
Capture WF
This function should only be used when instructed by a
Pulsar Measurement representative to do so. The
function captures the ultrasonic signal so that it can be
evaluated by Pulsar Measurement.
Select
Yes
to start the waveform download process.
After pressing
Yes
, the screen will flash
Working
for approximately 20 seconds, until the message
Insrt USB
appears. When
Insrt USB
is on
the screen, connect a flash drive to the USB port on the
front of the meter. The screen will flash
Saving
for
a couple seconds, and then return to
Done
. The
waveform is now stored on your flash drive and ready
to be sent to Pulsar Measurement.
Restore
Defaults
Select
Yes
to erase all user settings and return the
instrument to factory default settings. Note: does not
reset factory calibration values.
New Password
Select any number from 0000 to 9999. Default setting
of 0000 will allow direct access to the calibration
menus. Setting any password other than 0000 will
require the password to be entered to access the
calibration menus.
Press
to return to
Menu Selections
.
--Special Functions-
Language
English
Analog Out
4-20mA
Backlight
High
Reset Totalizer
NO
Neg. Totals
NO
Rev. Flow
OFF
Capture Par
NO
Capture WF
NO
Restore Defaults
NO
New Password
0000