![Pulsar Measurement Greyline TTFM 6.1 Instruction Manual Download Page 38](http://html1.mh-extra.com/html/pulsar-measurement/greyline-ttfm-6-1/greyline-ttfm-6-1_instruction-manual_1625415038.webp)
TTFM 6.1 INSTRUCTION MANUAL
38
2 or 4 CROSS INSTALLATION OVERVIEW – SE16A Transducers TMK-A1 Kit (Cont.)
4. Apply a small amount of coupling compound on the second transducer and place this transducer at the
separation distance provided in the Setup menu of the TTFM. Tighten this transducer down using the
built-in tightening bracket. Do not over-tighten the screw.
5. Coupling compound only: If you need to make fine adjustments (±0.1”) to the spacing at this point, you
may do so by loosening the tightening screw slightly, sliding the second transducer, then re-tightening it.