PULSAR MEASUREMENT
23
COMMUNICATION (Optional) (cont.)
Protocol
Choose
MODBUS
or
HART
.
Address
(Modbus)
Device address for the TTFM. Valid range: 001-247
(Default: 001). This number should be unique across the
bus. Press
or
to scroll,
to select digits, and
press
to store the setting.
BPS
(Modbus)
Baud rate for the MODBUS communications. Press
or
to select, and
to store the setting. Options:
4800, 9600, 19200, 38400, 57600, 76800, and 115200
(Default: 9600).
Parity
(Modbus)
Error checking parity for the MODBUS communications.
Press
or
to select, and
to store the setting.
Options: None, Even, and Odd (Default: Even).
Stop Bits
(Modbus)
Press
or
to select, and
to store the setting.
Options: 1 or 2 (Default: 1).
Note:
The Modbus register table, and HART configuration instructions can
be found in separate TTFM 6.1 Serial Communications Manual.