2 OR 4 CROSS INSTALLATION OVERVIEW – SE16B Transducers TMK-B1 Kit (Cont.)
5. Measure the separation distance from the mark you created in step 4 and create a new mark on the pipe
at the separation distance. It may be useful to mark your tape measure (included with every meter) at the
separation distance point before marking the pipe. The marked pipe is shown below.
6. Move the non-permanent bracket to position at the mark you created at step 5 and tighten it completely.
Apply coupling compound to the transducers and install them in the brackets.
Tips for installing transducers:
a. Be sure the red knob on the brackets are loosened completely
b. Put the transducer into the bracket by ensuring the bottom of the transducer and the couplant
does not touch the pipe as you slide it in. (Hover)
c. With the transducer hovering, tighten the red knob on the bracket until tight. The transducer will
be level with the surface of the pipe, and no coupling will have moved.
(pictures of proper coupling application and finished installation on the next page)