1 CROSS INSTALLATION OVERVIEW – SE16B Transducers TMK-B1 Kit (Cont.)
2. Install one of the stainless steel mounting brackets on the pipe. This will be the stationary bracket not
being rotated to the opposite side of the pipe. For a horizontal pipe, position the bracket at 3 or 9 o’clock.
Hold the level up to the top of the bracket to ensure the angle is correct. For vertical pipes, the bracket
can be at any orientation about the pipe. Tip: Use a 5/16” nut driver to tighten the hose clamp. Install the
second bracket at approximately the distance specified by the Sens Space value in the Setup menu.
3. Use alignment bar to ensure that the brackets are parallel. Completion of steps 2 and 3 is shown below.