TTFM 6.1 INSTRUCTION MANUAL
58
1 CROSS INSTALLATION OVERVIEW – SE16B Transducers TMK-B21 or TMK-B22 Kit (Cont.)
8. Position the marked mylar back on the pipe, with the overlap marks on the center line of the transducer
bracket that will be rotated to the other side of the pipe. Be sure the mylar is parallel with the face of the
transducer. Use duct tape to hold the mylar together and to the pipe. While the mylar is in this position,
mark the opposite side of the pipe where the mylar is marked from step 7: