TTFM 6.1 INSTRUCTION MANUAL
16
SET UP (cont.)
Sens pace
After sensor, angle, fluid, and pipe material are defined, this
displays the calculated sensor spacing. Also called the
separation distance. The sensors will be set to this
dimension when installed on the pipe, as described later in
this manual.
Velocity
Displays the measured velocity after the sensors have been
connected at the specified separation distance.
Signal
Strength
Displays magnitude of signal being received by the
ultrasonic sensor. Should be 100% under ideal operating
conditions. Signal strengths less than 100% do not indicate
that the meter is not reliable, however, the meter may be
more susceptible to complete signal loss should process
conditions like entrapped air worsen. When signal strength
is less than 100%, consideration should be made to using 1-
cross mounting method if this is not the current mounting
mode.
--
-------
Setup------
r
SE16B
Senso
Angle
37
°
Fluid
Water
Temp Mode
Fixed
Temp
77.0 F
Pipe
PVC
OD
4.5000 in
Wall
0.2500 in
Lining
None
Crossings
2
Zero Tare
No
Sens Space
2.299in
Velocity
0.00 ft/s
Signal
Strength
100
%