SEPARATION DISTANCE (Sensor Spacing Distance)
Separation distance is automatically calculated by the TTFM 6.1 based on parameters entered in the
Setup
menu.
Sens Space is a parameter where this distance is given, and it is located in the
Setup
menu. Document this value
for the following transducer installation procedure.
2 OR 4 CROSS INSTALLATION OVERVIEW – SE16B TMK-B1 Kit
1. Prep the pipe per instructions on page 29 and mind the installation location requirements on page 32.
Clean the location where the sensor is to be mounted on the opposite side of the pipe after we’ve marked
where it will be installed. Picture below shows a very clean ductile iron pipe which did not require much
cleaning. The outside paint is very well bonded and did not need to be removed:
2.
Install the stainless-steel mounting brackets on the pipe. Position them at approximately the correct
separation distance. Exact measurement is not required at this time. Tip: Use a 5/16” nut driver to tighten
the hose clamps.
Procedure continued on the next page…