1 CROSS INSTALLATION OVERVIEW – SE16B Transducers TMK-B21 or TMK-B22 Kit (Cont.)
9. Sand the pipe at this position if it needs to be prepped because of scale or rust. After prepped, move the
bracket to this 1 cross mark, and tighten it in place. Apply coupling compound to the transducers and
place them in the brackets:
View from front of pipe:
View from back of pipe:
Tips for installing transducers:
a. Be sure the tightening screws on the brackets are loosened completely.
b. Put the transducer into the bracket by ensuring the bottom of the transducer and the couplant
does not touch the pipe as you slide it in (hover).