15
© 2017 Proceq SA
sition is not optimal and the cover / diameter measurement becomes
not reliable
• Now move the probe cart horizontally until one of the Arrow LEDs F
lights up and then move back until MC LED C or D lights up.
• At this position you may also measure the diameter either by pushing
B on the right side of the probe or 5 (see Figure 7) on the Touchscreen
unit (e. g. when the probe is fixed to the telescopic extension rod).
• If the spacing of parallel or orthogonal rebars is between 5 and 30 cm
(2.00 to 12.00 inch), activate AI or NRC correction. If the cover is too
small for diameter measurement “too close” is displayed.
• In this case place a wooden or plastic board on the surface and set the
board thickness as Cover Offset to measure the diameter.
Finally the measured diameter must be set. The cover reading will be cor-
rected according to the diameter set.
NOTE! A rebar is shown only within the cover ranges indicat-
ed in Figure 3; for more details about diameter measurement
see “3.2.4. Practical Hints”.
• Tap
6
to store the diameter and cover measurement.
• Repeat this procedure at each rebar.
The saved data can be seen in the Snapshot and Normal Statistics View
(see “3.2.3. Cover Meter Display Views”).
Whenever a rebar is inclined to the Cover of the probe, if Display Inclined
Rebars is set, the angle will be shown on the display (minimum detected
inclination is 6 degrees).
NOTE! Cover values are only saved for later viewing, if the
diameter was measured and saved.
Single-Line View
Profometer
630 AI
Profometer
650 AI
1
2
3
4
5
6
7
8
9
10
16
17
14
15
13
12
11
19
18
Figure 8:
Single-Line Mode
1
File name:
enter the file name and tap return. Saved measurements
will be stored with this file name. If several measurements are made
under the same filename, a suffix increments after each measure-
ment and follows the file name.
2
Probe connected, current time, battery status and warning triangle
for zeroing the probe when needed: after 5 minutes in orange, after
10 minutes in red..
NOTE! Tap on the triangle to perform zeroing.