© 2017 Proceq SA
12
Return to start on new line
With this feature set, the cursor jumps back to the start line when chang-
ing line in the Multi-Line and Cross-Line Modes.
Line Height (in Y-direction)
The line height of must be set in the Multi-Line, Area-Scan and Cross-
Line Modes. It determines the spacing between the measuring rows. A
height of 5 to 203 cm / 2.00 to 80.00 inch can be set.
Grid Width (in X-direction)
The grid width must be set in the Area-Scan and Cross-Line Modes. A
width from 5 to 203 cm / 2.00 to 80.00 inch can be set.
3.2.2. Cover Meter Measurement Modes
When Measurement Mode is selected for the first time after switching on
the instrument, zeroing of the probe is performed. Confirm it and wait for the
button assessment window. Wait or tap anywhere on the screen. The Meas-
urement Modes available are shown in the menu on the top of the screen.
Locating
Single-Line
Multi-Line
Area-Scan
Cross-Line
Profometer 600
•
Profometer 630 AI
•
•
•
•
Profometer 650 AI
•
•
•
•
•
NOTE! Valid for all Measuring Modes: In case measuring data
shall be stored create a folder in the Explorer (see “5. Explorer
Document Handling”) and check if the correct folder is active.
The Maximum Cover should also be set for different files measured on
the same surface to get the same color range for comparison purposes.
Cover Offset
When a Cover Offset value is set the measured cover will be reduced by
this value; e. g. when a wooden or plastic plate is used to measure with
the probe cart on rough surfaces (see “3.2.4. Practical Hints”). In this
case the plate thickness must be set as Cover Offset value). A value from
1 to 50 mm / 0.04 to 1.92 inch can be set.
Display Inclined Rebar
By setting this feature the inclined rebar is displayed in the Locating Mode
when all four wheels of the cart have passed over the inclined rebar. In
the Single-Line and Multi-Line Modes it is only shown in the cart symbol.
NOTE! In areas with small rebar spacing this feature may not
work properly.
Sharpen
With this setting the signal strength color spectrum of the Multi-Line and
Cross-Line Views can be sharpened.
Display Curve
Select
Cover Value,
Signal Strength
or
None.
The respective curve or
no curve is displayed in Single-Line View.
Align Rebar Positions
When measuring in the Multi-Line or Cross-Line Mode along at least two
lines of at least 55 cm / 22.00 inch length, the rebar positions of subse-
quent lines are aligned to the rebar positions of the previous lines.
NOTE! This feature should only be set if the rebars are running
parallel to the start line (X- or Y-line). It is not activated during
the measurements (activated only when storing the data).