23
© 2017 Proceq SA
Snapshot View
Profometer
600
Profometer
630 AI
Profometer
650 AI
The Snapshot View can be displayed if at least one diameter was meas-
ured and stored in one of the measurement Modes.
The cover values are shown as vertical bars to scale and the diameter as
a figure, both in the unit set. The Minimum Cover is not displayed in the
Snapshot View.
Figure 17:
Snapshot View
Statistics View
Profometer
600
Profometer
630 AI
Profometer
650 AI
The Statistics View can be displayed for measurements done and stored
in one of the measurement Modes. It shows the statistical calculation of
the cover values measured.
For measurements with the Cross-Line Mode the statistical evaluation of
the cover readings is done for each layer independently. Hence there is a
Statistical View each for the scanning in the X- and Y-direction.
NOTE! In practice only the cover values and statistical evalu-
ation of the 1st layer rebars (closer to the surface) is of inter-
est.
On the horizontal axis the cover values in the unit set are displayed. The
vertical bars show the percentage of the respective cover values meas-
ured and stored. The vertical cursor bar can be moved to any cover value.
The figure on the left of the cursor bar shows the percentage of measured
cover values smaller than the cursor position. The value on the right shows
percentage of measured cover values larger than the cursor position. The
cover value is displayed at the bottom of the cursor bar and at the top the
percentage of measured covers for that cover is shown. Minimum required
cover is shown as a vertical dotted line in red (if set). Covers below the min-
imum are shown as red bars, covers above the minimum as yellow bars.
There are two different Statistics Views, the
Normal
(see Figure 18) and
the
DBV-Evaluation
(see Figure 19). Tap on the statistical values win-
dows to switch from Normal to DBV.
Statistics values box “Normal” showing Median, Mean, Number of covers measured, lowest/
highest, Standard Deviation.
Change from X- to Y-direction view (for Cross-Line Mode data only) by tapping on
/
.
The actual window is shown on top right (either of scan direction SX or SY)
Figure 18:
Statistics View Normal