82
Revision 00 EN - 02/2022
MSM+ HB
35,0(6
Option
Explanation
TriggerLevel *
This drop-down list is only visible if the drop-down area for displaying the
options of the trigger operation is open.
The trigger level is the threshold value for triggering a signal via the inter-
nal photodiode. Use one of the following options to set the trigger level:
X
Enter a value in the input field.
X
Use the slider below the input field.
You can determine the threshold value suitable for a measurement task
as follows:
1. Set the trigger level to the maximum value.
2. Decrease the value step by step until the device receives some
trigger signals (lower trigger level).
3. Increase the value until the device does not receive any trigger
signals (upper trigger level).
4. Calculate the trigger level as the arithmetic mean of the two limit
values.
Locked area
Locked area means a restriction of the traversing range in the z-direc-
tion. The lower limit of the enabled traversing range is always home
position. The upper limit can be set freely.
Maximum z-Position in mm *
1. Use one of the following options to set the locked area:
X
Enter the upper limit of the permitted z-traversing range in the input
field.
X
Use the slider below the input field.
We recommend saving a locked area in the device, if it is safety-relevant:
2. If you want to save the locked area (and the other options marked
with an asterisk) in the device, click the
Save device settings
button.
The settings are retained even if the device is switched off or dis-
connected from the power source.
Tab. 9.10: Options in the
Device control > Advanced
tab of the
Automatic caustic
measuring mode
Summary of Contents for MicroSpotMonitor Plus HighBrilliance
Page 2: ......
Page 3: ...3 Revision 00 EN 02 2022 MSM HB 35 0 6 IMPORTANT READ CAREFULLY BEFORE USE KEEP FOR FUTURE USE...
Page 7: ...7 Revision 00 EN 02 2022 MSM HB 35 0 6...
Page 95: ...95 Revision 00 EN 02 2022 MSM HB 35 0 6 13 Declaration of conformity...
Page 96: ...96 Revision 00 EN 02 2022 MSM HB 35 0 6...
Page 98: ...98 Revision 00 EN 02 2022 MSM HB 35 0 6 15 Dimensions 15 1 Dimensions of the MSM HB10 A...
Page 99: ...99 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB10 continued View A...
Page 100: ...100 Revision 00 EN 02 2022 MSM HB 35 0 6 15 2 Dimensions of the MSM HB10 with fibre bridge A...
Page 103: ...103 Revision 00 EN 02 2022 MSM HB 35 0 6 15 3 Dimensions of the MSM HB20 at zmax C...
Page 104: ...104 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB20 continued View C...