16.3.2
Measuring configuration selection
There are various measuring options to differentiate between:
• Measuring a single plane or a caustic
• Measuring a complete pulse or just a single section
• Measuring with freely adjustable integration time or with automatic integration time control
• Measuring in trigger operation or quasi-cw operation
• Variations of optimal integration time caused by changing the attenuation
Combining these measurement options with the following pulse parameters provides numerous possibilities:
• Pulse duration: fs – ms
• Pulse frequency: 1 Hz – 1 kHz
The following merely describes a rough structure that is intended to help in choosing measuring settings.
16.3.3
Influence of the pulse parameters on the integration time control
The software-operated integration time control always assumes that there is a cw laser beam.
This may cause quantization of the integration time for slow pulse lasers (< 500 Hz) or lasers with high pulse
energy (integration time very short). Tab. 16.1 And the diagram in Fig. 16.3 on page 112 illustrates this.
Pulse frequency
Number of detected pulses
in Hz
at maximum integration time 225 ms
at integration time 1 ms
1
0
0 – 1
5
1
0 – 1
10
2
0 – 1
50
11
0 – 1
100
22
0 – 1
200
45
0 – 1
500
112
0 – 1
1 000
225
1 – 2
2 000
450
2 – 3
5 000
1 125
5.00
10 000
2 250
10.00
Tab. 16.1: Number of detected pulses in correlation with the integration time and pulse frequency
Summary of Contents for MicroSpotMonitor Plus HighBrilliance
Page 2: ......
Page 3: ...3 Revision 00 EN 02 2022 MSM HB 35 0 6 IMPORTANT READ CAREFULLY BEFORE USE KEEP FOR FUTURE USE...
Page 7: ...7 Revision 00 EN 02 2022 MSM HB 35 0 6...
Page 95: ...95 Revision 00 EN 02 2022 MSM HB 35 0 6 13 Declaration of conformity...
Page 96: ...96 Revision 00 EN 02 2022 MSM HB 35 0 6...
Page 98: ...98 Revision 00 EN 02 2022 MSM HB 35 0 6 15 Dimensions 15 1 Dimensions of the MSM HB10 A...
Page 99: ...99 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB10 continued View A...
Page 100: ...100 Revision 00 EN 02 2022 MSM HB 35 0 6 15 2 Dimensions of the MSM HB10 with fibre bridge A...
Page 103: ...103 Revision 00 EN 02 2022 MSM HB 35 0 6 15 3 Dimensions of the MSM HB20 at zmax C...
Page 104: ...104 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB20 continued View C...