64
Revision 00 EN - 02/2022
MSM+ HB
35,0(6
During the measurement, the pro-
gress is shown in the following indica-
tors:
Measuring plane
While the indicator is rotating, the
measurement is performed.
Measurement completed
After successful measurement, the indica-
tor lights up green.
10. Switch off the laser after the
measurement is completed, unless
you want to perform further measu-
rements.
Averaging (if enabled)
The indication shows the measured pla-
nes that are used to average a measured
value.
Adjusting size and position of the
measurement window (optional)
11. Make sure that the
autom.
Measurement window
option is not
enabled. Otherwise, the manual set-
ting may be overwritten when starting
a measurement.
12. Use one of the following options to
adjust the size of the measurement
window:
X
Enter the length and width in the
corresponding fields.
X
Position the mouse pointer anywhe-
re within the measurement range
and drag while holding down the
left mouse button. Keep the mouse
button pressed until the measurement
window meets your requirements.
13. Use one of the following op-
tions to adjust the position of the
measurement window:
X
Enter the x-position / y-position of the
center of the measurement window in
the corresponding fields.
X
Position the mouse pointer within the
measurement window. Then drag
while holding down the right mouse
button.
14. Click the Start button to perform the
measurement in the manually custo-
mized measurement window.
15. Switch off the laser after the
measurement is completed, unless
you want to perform further measu-
rements.
Summary of Contents for MicroSpotMonitor Plus HighBrilliance
Page 2: ......
Page 3: ...3 Revision 00 EN 02 2022 MSM HB 35 0 6 IMPORTANT READ CAREFULLY BEFORE USE KEEP FOR FUTURE USE...
Page 7: ...7 Revision 00 EN 02 2022 MSM HB 35 0 6...
Page 95: ...95 Revision 00 EN 02 2022 MSM HB 35 0 6 13 Declaration of conformity...
Page 96: ...96 Revision 00 EN 02 2022 MSM HB 35 0 6...
Page 98: ...98 Revision 00 EN 02 2022 MSM HB 35 0 6 15 Dimensions 15 1 Dimensions of the MSM HB10 A...
Page 99: ...99 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB10 continued View A...
Page 100: ...100 Revision 00 EN 02 2022 MSM HB 35 0 6 15 2 Dimensions of the MSM HB10 with fibre bridge A...
Page 103: ...103 Revision 00 EN 02 2022 MSM HB 35 0 6 15 3 Dimensions of the MSM HB20 at zmax C...
Page 104: ...104 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB20 continued View C...