Revision 00 EN - 02/2022
MSM+ HB
35,0(6
cw operation
Pulse frequency
Pulse duration
1 μs
1 Hz
1 MHz
1 kHz
10 Hz
10 kHz
100 Hz
100 kHz
1 s
10 μs
10 s
100 μs
100 s
Trigger operation
Quasi-cw operation
Fig. 16.5:
Selection of the appropriate measurement mode based on laser parameters
For the blue field laser beams, the quasi-cw operation is most suitable. However, be aware that as the
pulse frequency decreases, the required integration time increases to ensure the quasi-cw case. As a rule
of thumb, the integration time of a focal plane measurement should be approximately the same as the time
for 35 pulses. If the laser beam to be measured falls below a pulse frequency of approx. 500 Hz, you should
switch to trigger operation.
While it is almost always possible to measure with the integration time control in the cw or quasi-cw ope-
ration, it only makes sense to use it for very long pulse durations (>1 ms) in trigger operation. With the help
of the attenuation filters, the integration time is adjusted in such a way that it is only a fraction of the pulse
duration. The trigger signal will then merely specify to the device the starting time for the measurement. The
integration time may increase or decrease during the course of the caustic measurement without leaving the
pulse path (see Fig. 16.6 and “Example 2: very long pulse duration” on page 115).
Integrationszeit
Integration time (e. g. 100 μs)
P
Laser
t
min Delay (12 μs)
tp – 1 ms
Fig. 16.6:
Measuring parameters for pulsed laser systems with pulse duration greater than 1 ms
For all other cases, it is advisable to specify a fixed integration time in order to always measure a fixed num-
ber of pulses by cleverly selecting the filters and the delay and integration time (“Example 1: fixed integration
time” on page 115).
Summary of Contents for MicroSpotMonitor Plus HighBrilliance
Page 2: ......
Page 3: ...3 Revision 00 EN 02 2022 MSM HB 35 0 6 IMPORTANT READ CAREFULLY BEFORE USE KEEP FOR FUTURE USE...
Page 7: ...7 Revision 00 EN 02 2022 MSM HB 35 0 6...
Page 95: ...95 Revision 00 EN 02 2022 MSM HB 35 0 6 13 Declaration of conformity...
Page 96: ...96 Revision 00 EN 02 2022 MSM HB 35 0 6...
Page 98: ...98 Revision 00 EN 02 2022 MSM HB 35 0 6 15 Dimensions 15 1 Dimensions of the MSM HB10 A...
Page 99: ...99 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB10 continued View A...
Page 100: ...100 Revision 00 EN 02 2022 MSM HB 35 0 6 15 2 Dimensions of the MSM HB10 with fibre bridge A...
Page 103: ...103 Revision 00 EN 02 2022 MSM HB 35 0 6 15 3 Dimensions of the MSM HB20 at zmax C...
Page 104: ...104 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB20 continued View C...