80
Revision 00 EN - 02/2022
MSM+ HB
35,0(6
Option
Explanation
Filter level
The selected setting will only be active if the
autom. filter adjustment
option is disabled.
The filter required for measurement depends on the wavelength and
intensity of the laser beam to be measured. We recommend selecting
a filter level that achieves a detector modulation of between 60 % and
80%.
X
Select a filter level in the drop-down list.
autom. filter adjustment
If this option is enabled, then the filter levels are set automatically.
X
Set the check mark to enable the option.
Time series
A time series consists of several caustic measurements with the same
settings.
Number of measurements *
X
Enter the desired number of caustic measurements of the time
series.
Interval between measurements in s *
X
Enter the pause between the caustic measurements of the time
series. This is the time gap between the end of one measurement
and the start of the next.
Tab. 9.9:
Options in the
Device control > Settings
tab of the
Automatic caustic
measuring mode
9.8.2
Configure advanced settings
1. Click on the
Advanced
tab.
2. Maintain the options according to the
explanations in the following table..
Option
Explanation
Save device settings
All options marked with an asterisk in the
Device control
menu can be
saved in the EEPROM of the device.
These and other options for saving / loading configurations are descri-
bed in chapter 9.4.4 on page 53.
# Pixel in x/y *
This option enables setting the resolution of the CCD sensor:
X
Enter the number of pixels in the x-direction / y-direction in the input
field on the left / right.
Number of planes *
The default setting is 21 planes.
X
Enter the number of planes to be measured.
Low power mode
We recommend enabling this option for relatively low power laser be-
ams.
If this option is enabled, the following initial values are selected during
beam search
• Filter level of the filter wheel: 0 (instead of 1 otherwise)
• Integration time: 10 ms (instead of 1 ms otherwise).
X
Set the check mark to enable the option.
Tab. 9.10: Options in the
Device control > Advanced
tab of the
Automatic caustic
measuring mode
Summary of Contents for MicroSpotMonitor Plus HighBrilliance
Page 2: ......
Page 3: ...3 Revision 00 EN 02 2022 MSM HB 35 0 6 IMPORTANT READ CAREFULLY BEFORE USE KEEP FOR FUTURE USE...
Page 7: ...7 Revision 00 EN 02 2022 MSM HB 35 0 6...
Page 95: ...95 Revision 00 EN 02 2022 MSM HB 35 0 6 13 Declaration of conformity...
Page 96: ...96 Revision 00 EN 02 2022 MSM HB 35 0 6...
Page 98: ...98 Revision 00 EN 02 2022 MSM HB 35 0 6 15 Dimensions 15 1 Dimensions of the MSM HB10 A...
Page 99: ...99 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB10 continued View A...
Page 100: ...100 Revision 00 EN 02 2022 MSM HB 35 0 6 15 2 Dimensions of the MSM HB10 with fibre bridge A...
Page 103: ...103 Revision 00 EN 02 2022 MSM HB 35 0 6 15 3 Dimensions of the MSM HB20 at zmax C...
Page 104: ...104 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB20 continued View C...