78
Revision 00 EN - 02/2022
MSM+ HB
35,0(6
9.7.3
Measure a caustic manually
DANGER
Serious eye or skin injury due to laser radiation
X
Observe the warning messages in chapter 9.1 on page 44.
1. Switch on the laser.
2. Click the
Start
button.
The measurement begins.
3. Optional:
If necessary, click the
Stop
button to
end the measurement prematurely.
During the measurement, the pro-
gress is shown in the following indica-
tors:
Measuring caustic
While the indicator is rotating, the
measurement is performed. The
measurement parameters are optimized
according to Primes quality criteria.
Measurement completed
After successful measurement, the indica-
tor lights up green.
4. Switch off the laser after the
measurement is completed, unless
you want to perform further measu-
rements.
Averaging (if enabled)
The indication shows the measured pla-
nes that are used to average a measured
value.
Applying a caustic suggestion (optio-
nal)
After the measurement is completed,
a z-range is displayed in the
Caustic
suggestion
fields. The limits of the
suggested z-range have a distance of
± 3 Rayleigh lengths to the determi-
ned beam waist position.
This z-range is optimized according
to Primes quality criteria.
5. If you want to apply this suggestion,
click the
Use
button.
The suggested limit values z1 / z2 are
transfered to the
z-Axis position (z1 /
z2) in mm
fields.
6. Click the
Start
button again to per-
form the measurement in the sugges-
ted z-range.
7. Switch off the laser after the
measurement is completed, unless
you want to perform further measu-
rements.
Summary of Contents for MicroSpotMonitor Plus HighBrilliance
Page 2: ......
Page 3: ...3 Revision 00 EN 02 2022 MSM HB 35 0 6 IMPORTANT READ CAREFULLY BEFORE USE KEEP FOR FUTURE USE...
Page 7: ...7 Revision 00 EN 02 2022 MSM HB 35 0 6...
Page 95: ...95 Revision 00 EN 02 2022 MSM HB 35 0 6 13 Declaration of conformity...
Page 96: ...96 Revision 00 EN 02 2022 MSM HB 35 0 6...
Page 98: ...98 Revision 00 EN 02 2022 MSM HB 35 0 6 15 Dimensions 15 1 Dimensions of the MSM HB10 A...
Page 99: ...99 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB10 continued View A...
Page 100: ...100 Revision 00 EN 02 2022 MSM HB 35 0 6 15 2 Dimensions of the MSM HB10 with fibre bridge A...
Page 103: ...103 Revision 00 EN 02 2022 MSM HB 35 0 6 15 3 Dimensions of the MSM HB20 at zmax C...
Page 104: ...104 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB20 continued View C...