PDC 2000/3000 Repair Manual
Testing&Adjustment
7. Set the 1-ohm resistor box Manual/LV switch to LV (LabView).
8. Click on the white arrow. Arrow should change to black.
9. Click on the black arrow to start the selected LabView VI test program.
The test program starts as indicated by the yellow Running box directly under the black
start arrow.
Note: Follow all instructions that appear on the screen while the test program runs.
Click on the appropriate responses that appear in the dialog boxes.
10. Upon completion of the current profile test program, current waveforms for the following
appear:
•
Current (Amperes)
-
Waveform should be smooth.
•
Reference Current (Amperes)
-
Waveform should be smooth.
•
Difference Current (Amperes)
-
Not used.
•
Filtered Windowed Difference current (Amperes)
-
Not used.
Note: Spikes displayed in the current profile waveforms indicate that the camera is
defective. Refer to the camera troubleshooting section in the repair manual to
determine the cause of the camera malfunction and how to correct it.