- 78 -
3.3.b Too Low
Press the y-top limit up and y-bottom limit up. (Figure 143)
Figure 143
Press
Test
and take another exposure. If the beam is aligned then press
Done
. Move onto the Secondary collimator (beaver tail).
4
Ceph Head Leveling
Put the left and right ear posts into their respective holders. (Figure 144)
Figure 144
Printed copies of this document are considered uncontrolled.
22811.20.Rev001 07.31.2018
78
Summary of Contents for ProMax Pan/Ceph
Page 6: ...6 Printed copies of this document are considered uncontrolled 22811 20 Rev001 07 31 2018 6...
Page 8: ...8 Printed copies of this document are considered uncontrolled 22811 20 Rev001 07 31 2018 8...
Page 16: ...16 Printed copies of this document are considered uncontrolled 22811 20 Rev001 07 31 2018 16...
Page 18: ...18 Printed copies of this document are considered uncontrolled 22811 20 Rev001 07 31 2018 18...
Page 38: ...38 Printed copies of this document are considered uncontrolled 22811 20 Rev001 07 31 2018 38...