3.2
Ceph mode beam position (61.6)
Press
Ceph mode beam position
. (Figure 135)
Figure 135
Press
Test
and take an exposure. Beam should cover the center of the
fluorescent screen with a bit of overlap on either side of the center lines.
(Figure 136)
Figure 136
Adjust if the beam is off.