4.3 Too Low
If the ball is too low then adjust arm at points in Figure 174.
Figure 174
Loosen #1 a quarter of a turn then loosen up #3 and tighten down #2.
Take another exposure. If aligned then tighten #1 a quarter turn. If aligned, re-
check Beam check.
4.4
Too High
If the ball is too high then adjust arm at points in Figure 175.
Figure 175
Loosen #1 a quarter of a turn then loosen up #2 and tighten down #3.
Take another exposure. If aligned then tighten #1 a quarter turn. If aligned, re-
check Beam check
Printed copies of this document are considered uncontrolled.
22811.20.Rev001 07.31.2018