- 95 -
Take an exposure. The image should be the same on the top, bottom, left, and
right. (Figure 199)
Figure 199
7
Calibrating
Press
Exit
on the touch screen, if just finished doing a Beam Check.
7.1 kV / mA for Cephalometric
Binnings
kV
mA
4x4
60
4
3x3
60
7
Printed copies of this document are considered uncontrolled.
22811.20.Rev001 07.31.2018