- 26 -
2.1.a.ii Too Wide
Press the X-pan collimator width to the left (Figure 33)
Figure 33
Press
Test
and take another exposure.
If aligned then press
Done
and move onto the Y
Collimator.
2.2 Pan mode Y Collimator (61.2)
Now check the top and bottom of the beam.
2.2.a Too High
Press the Y-pan collimator top limit button down. (Figure 34)
Figure 34
Printed copies of this document are considered uncontrolled.
22811.20.Rev001 07.31.2018
26