- 54 -
3
Position Sensor Calibration (Pots)
NOTE: If the ball phantom image is good, then skip this section.
Press
i
then select
Technical Calibration
(60.0) then select
Position Sensor
Calibration.
(64.0). (Figure 94)
Figure 94
Press
Position Sensor
(64.1) (Figure 95)
Figure 95
Printed copies of this document are considered uncontrolled.
22811.20.Rev001 07.31.2018
54