- 57 -
Watch the numbers on the touch screen and rotate spindle until in the
Green and center. (Figure 100)
Figure 100
Put sensor coupling back onto spindle and re-tighten set screw. Re-
check still in the green and center.
When aligned take out the pins then press
Done
.
3.1.b Elbow
Unscrew the set screw with a 2mm allen to get the sensor holder off of
the cable guide. (Figure 101)
Figure 101
Printed copies of this document are considered uncontrolled.