Loosen the 3mm locking screw on the outside, closet to the column. (Figure
189)
Figure 189
Take another exposure. If aligned then re-tighten the inner screws and move
onto Beam Check.
5.5 Angled
5.5.a Left
If angled left, then loosen the nuts and loosen the right locking screw
and tighten the left locking screw. (Figure 190)
Figure 190
Take another exposure. If aligned then re-tighten the nuts and move
onto Beam Check.