- 30 -
The collimator will move to expose one screw then push the lower blade up
and expose the second screw. (Figure 41)
Figure 41
NOTE: This allows the two screws that adjust the angle to be visible.
2.2.c.i Angled Left/Right
Loosen the two hidden screws with a 2.5 mm allen at # 1.
(Figure 42)
Figure 42
Twist collimator to the right if angled Right to Left
Twist collimator to the left if angled Left to Right
Tighten down the screws and take another exposure.
Go back into Pan X-Collimator and Y-Collimator and press
Test
in
both allowing the collimator to go to the home position.
Take another exposure. Press
Done
until you get back to the Technical
Calibration Screen.
Printed copies of this document are considered uncontrolled.
22811.20.Rev001 07.31.2018
30