- 84 -
5
Secondary Collimator (Beaver Tail)
Unscrew the three 3mm allen screws. (Figure 176a-b)
Figure 176a
Figure 176b
Attach the collimator and screw in the three 3mm allens. (Figure 177a-b)
Figure 177a
Figure 177b
Press
i
then
Technical calibration
then
Primary Collimator
and select “Ceph mode
beam position” (61.6)
Press
Test
and take an exposure. (Figure 178)
Figure 178
Printed copies of this document are considered uncontrolled.