- 48 -
NOTE: The new shoulder number should be lower than old number.
Press
Lock
and take the pin out and then press
Done
. Take another exposure
and check ball size again. If not repeat until correct size.
2.1.b 10
th
Ball
Measure from the center of the Center Ball to the center on the 10
th
Ball on
the right then measure from the center of the Center Ball to the center of 10
th
Ball on the left. (Figure 82)
Figure 82
NOTE: From the Center Ball, the distance left or right to the 10
th
Ball will always need to be the
same.
2.1.b.i
Too Left
If the distance from the Center Ball to the left is more than the
right, then move the table away from the column. (Figure 83).
Figure 83
Printed copies of this document are considered uncontrolled.