- 9 -
2
REQUIRED TOOLS
2.1.
Calibration Tools
Ball phantom (part number 10004011) used for checking the position of the
Patient positioning mechanism and the Patient positioning lights. (Figure 1)
Figure 1
Frankfort plane alignment tool (part number 50977) used with the ball
phantom for checking the position of the Frankfort plane light. (Figure 2)
Figure 2
Beam alignment tool, fluorescent screen, (part number 50972) used for
checking the position of the x-ray beam. (Figure 3)
Figure 3
Printed copies of this document are considered uncontrolled.
22811.20.Rev001 07.31.2018
9