- 89 -
Tighten the 3mm locking screw on the outside, closest to the column. (Figure
187)
Figure 187
Take another exposure. If aligned then retighten the inner screws and move
onto Beam Check.
5.4 Too High
Loosen the two 3mm allen screws on the inside of the collimator. (Figure 188)
Figure 188