- 43 -
2.1.a Center Ball
NOTE: Center ball will be above the midsagittal bar of the ball phantom. (Figure 67)
Figure 67
Click the second icon now and measure the top to bottom and left to
right. (Figure 68)
Figure 68
The ball should be 7mm top to bottom and left to right. (Figure 69)
Figure 69
NOTE: If ball phantom is off, do a Position Sensor Calibration (Pots) on pgs. 49-54.