© 2020 - perma-tec
GmbH & Co. KG
|
www.perma-tec.com
|
27
26
|
perma ECOSY
Parameter
Vol. outlet 3
0000 ml/1000h
Edit Esc
B6
The currently set discharge volume of
the selected outlet is displayed.
To change, press “EDIT” (after you have
confirmed the selected PIN according
to chapter 8.4.6).
B7
Parameter
Vol. outlet 3
0000 0000
OK -- ++ >>
B7
The first digit of the value to be changed
is underlined and blinking.
With “++” or “--”, you can change the
value for this digit (no changes in this
example) or go to the next digit with
“>>”.
B8
Parameter
Vol. outlet 3
0000 0000
OK -- ++ >>
B8
The selected digit is underlined and
blinking (here, the second 0).
With “++” or “--”, you can set the
required value for this digit (here,
the value was increased to 5 for 500).
B9
Parameter
Vol. outlet 3
0000 0500
OK << >> Esc
B9
Go to the next digit with “>>”.
Continue as described above until
you have completed setting all desired
values.
B10
Parameter
Vol. outlet 3
0000 0500
OK << >> Esc
B10
Save the displayed value with “OK”.
B11
Parameter
Vol. outlet 3
0500 ml/1000h
Edit Esc
B11
Exit this menu with “ESC”.
Every time you press “ESC”, you will
reach a higher menu level.
Tightly seal all inactive outlets with a plug (see menu “parameter” outlets
set to 0 discharge volume).