Installing Your RamanStation 400 .
49
Attaching the Fiber Port Cover Assembly
NOTE: This section is only applicable if you have a RamanStation 400F.
The RamanStation 400F is classified as a Class 3B laser product according to the label
attached to the instrument. However if required, the RamanStation 400F can be classified as
a Class 1 laser product provided the following conditions are met:
•
No fiber optic probe is connected to the instrument;
•
The fiber optic probe cover assembly (L1320263) is attached to the instrument and
completely encloses the EXC and COL couplers on the rear panel.
A Class 1 laser product is safe under all reasonably foreseeable conditions of use and may be
used in an open laboratory environment.
To attach the fiber port cover to the rear panel, fasten the supplied screw through the cover
into the hole near the COL coupler as shown in Figure 17.
Figure 17 Attaching the fiber port cover
Screw
Summary of Contents for RamanStation 400
Page 1: ...RamanStation 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY...
Page 5: ...Introduction...
Page 10: ...10 RamanStation 400 Series Getting Started Guide...
Page 11: ...Safety Practices...
Page 28: ...28 RamanStation 400 Series Getting Started Guide...
Page 29: ...Raman History Theory and Instrumentation...
Page 37: ...Preparing Your Facility...
Page 42: ...42 RamanStation 400 Series Getting Started Guide...
Page 43: ...Installing Your RamanStation 400...
Page 71: ...Introduction to the RamanStation 400...
Page 76: ...76 RamanStation 400 Series Getting Started Guide...
Page 77: ...Fiber Optic Probe...
Page 84: ...84 RamanStation 400 Series Getting Started Guide...
Page 85: ...Using the RamanStation 400...
Page 106: ...106 RamanStation 400 Series Getting Started Guide...
Page 107: ...Maintenance and Troubleshooting...
Page 118: ...118 RamanStation 400 Series Getting Started Guide...