Attaching and Removing the Fiber Optic Probe
The standard fiber optic probe is constructed such that a single armored cable bifurcates into
an excitation cable and a collection cable. The excitation fiber transmits the laser energy,
whereas the collection cable is a conduit for the Raman energy. Both fibers terminate in FC
style connectors. There is also a plug attached to the end of two wires which connects the
fiber optic LED to the instrument.
Attaching the FOP
1.
Ensure the attenuator cap is attached to the FOP.
2.
Turn off the instrument and remove the main power cable and, if appropriate, the stage
controller power cable.
3.
Carefully move the instrument to permit easier access to the rear panel.
4.
Remove the fiber port cover assembly if fitted (see
Attaching the Fiber Port Cover
Assembly
on page 49).
5.
Remove the retained screw-in caps from the coupler area at the bottom left of the
panel.
6.
Connect the excitation cable of the FOP to the coupler marked EXC.
Make sure that the polarizing ridge on the FOP cable is aligned with the slot in the
coupler casing before carefully pushing the cable into the coupler (Figure 27).
Figure 27 Connecting the fiber optic probe
7.
Connect the collection cable of the FOP to the coupler marked COL.
Make sure that the polarizing ridge on the FOP cable is aligned with the slot in the
coupler casing before carefully pushing the cable into the coupler.
8.
Connect the plug to the nearby socket marked LED FIBER #1.
9.
Reconnect the power cables.
Coupler
Probe
Polarizing ridge
aligned with slot
Summary of Contents for RamanStation 400
Page 1: ...RamanStation 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY...
Page 5: ...Introduction...
Page 10: ...10 RamanStation 400 Series Getting Started Guide...
Page 11: ...Safety Practices...
Page 28: ...28 RamanStation 400 Series Getting Started Guide...
Page 29: ...Raman History Theory and Instrumentation...
Page 37: ...Preparing Your Facility...
Page 42: ...42 RamanStation 400 Series Getting Started Guide...
Page 43: ...Installing Your RamanStation 400...
Page 71: ...Introduction to the RamanStation 400...
Page 76: ...76 RamanStation 400 Series Getting Started Guide...
Page 77: ...Fiber Optic Probe...
Page 84: ...84 RamanStation 400 Series Getting Started Guide...
Page 85: ...Using the RamanStation 400...
Page 106: ...106 RamanStation 400 Series Getting Started Guide...
Page 107: ...Maintenance and Troubleshooting...
Page 118: ...118 RamanStation 400 Series Getting Started Guide...