Problem:
A motorized sample accessory does not go to the correct position
when asked.
Description This means that for some reason the motorized stage is jammed. This
typically happens if the sample holder on the stage is different to that
selected on the Sampling tab in the Scan and Instrument Setup dialog.
Solution 1
Go to the Sampling tab and ensure that the selected sample holder is
correct. If it is not then select the appropriate sample holder.
Solution 2
Go to the Stage Control section and then click
Reset Stage
. The motorized
accessory will now reset itself.
Summary of Contents for RamanStation 400
Page 1: ...RamanStation 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY...
Page 5: ...Introduction...
Page 10: ...10 RamanStation 400 Series Getting Started Guide...
Page 11: ...Safety Practices...
Page 28: ...28 RamanStation 400 Series Getting Started Guide...
Page 29: ...Raman History Theory and Instrumentation...
Page 37: ...Preparing Your Facility...
Page 42: ...42 RamanStation 400 Series Getting Started Guide...
Page 43: ...Installing Your RamanStation 400...
Page 71: ...Introduction to the RamanStation 400...
Page 76: ...76 RamanStation 400 Series Getting Started Guide...
Page 77: ...Fiber Optic Probe...
Page 84: ...84 RamanStation 400 Series Getting Started Guide...
Page 85: ...Using the RamanStation 400...
Page 106: ...106 RamanStation 400 Series Getting Started Guide...
Page 107: ...Maintenance and Troubleshooting...
Page 118: ...118 RamanStation 400 Series Getting Started Guide...